Oct 13, 2013
Mar 4, 2013
Tech Talk - Medical Device Particle Testing Part 3
- Perform test on water with glassware
- Perform test on water through the model without test device
- Perform test on water through the model after test device is cycled
Oct 27, 2011
Tech Talk – In Vitro Medical Device Verification Testing in Blood
I previously discussed testing medical devices in blood here (in 2007!), but I think I did a poor job of it and I’d like to revisit it.
Why do you test in blood? Well for one, blood is hard to simulate, it’s a non-Newtonian fluid, and using glycerin and water don’t really do it justice, but these can work depending on the application. For another, a common blood test is to check for hemolysis, sure this is tested during a biocompatibility test, but biocompatibility tests are not performed during actual use conditions. Hemolysis may also be a part of an animal safety study that you want to check out beforehand.
Where do you get the blood from? At a slaughterhouse of course, if you can find a smaller or craft meat location in your area, they’ll probably work with you, one used to sell to us for $40 a week, and we’d take a couple gallon buckets and their workers would fill them up while we waited. They only slaughtered on certain days, so call ahead. You’ll probably find cows easier to find and work with, but there isn’t really a reason you couldn’t use pig blood.
Before we leave for the slaughterhouse we’ll set up a water bath at 37ÂșC to be ready when we get back. Then we’ll add anticoagulant to the blood collection bucket. We’ll use either heparin or Acid Citrate Dextrose (ACD).
Once we get the blood, we mix the bucket to ensure the anticoagulant is distributed in the blood. Heparin is prescription drug, so hit up your vet consultant or animal lab for some ahead of time. ACD you can make based on USP guidelines from commonly available chemicals (water, citric acid, dextrose, and sodium). We used around 10,000 to 20,000 units of heparin per liter of blood. Of note is heparin is used clinically (on people) more in the U.S. and ACS is used in Europe, so you could maybe argue for the use of one over the other, but you’re using animal blood, so I’m not sure if that really matters. I’ll assume we are using bovine blood for the rest of this post. If you don’t use an anticoagulant, you’ll end up with a clot bucket when you get back to the lab, just throw it away if this happens, it is not recoverable.
Time is generally of the essence so don’t stop by Chili’s on your way back to the lab. Also, just be aware that water will damage your blood cells, so it is preferable to rinse your lab ware with a bit of saline before use.
When we get back to the lab we first check the blood pH and temperature, ideally the pH is between 7.2 and 7.4. We then take a hematocrit (hct) measurement by collecting blood in a capillary tube with clay sealant to stopper the bottom (get blood before using clay). Then we centrifuge the capillary tube for a few minutes at high rpm. Once centrifuged, the capillary tube will look like this:
You’ll need a hematocrit chart. Below is a simple representation of how to measure hematocrit, you put the capillary tube on the chart, line up the clay on the baseline, you move the tube left or right until the fluid level matches the top line, then you find the line where the red blood cells stop and follow it over to read the percent hematocrit, in this case 50%.
We generally take two hematocrit measurements and average; you need two capillary tubes to balance the centrifuge anyway. 38 to 42% hct is a typical range for a study like this one, although it will vary depending on how much the animal drank before it was slaughtered, I’ve seen it come in in the low 20s, so don’t worry about the initial hematocrit too much.
We then pump the blood from the collection bucket through a saline primed arterial filter (pediatric filters have lower priming volumes) and line to remove hair and large clots and into a carboy with a plugged outlet at the bottom. We’ll set up a circuit from the bottom of the carboy to the top (through the filter) with a peristaltic pump to keep the blood circulating. Place the outlet in the blood and not above it or you’ll get a bunch of foam. At this point we’ll add saline to the blood to get the hematocrit where we want it, usually around 22% to 32%. Keeping hematocrit consistent is better than not. We’ll measure the hematocrit and adjust until we’re good, using the following formula:
H is the initial hct,
F is the desired hct,
V is the original volume of blood, and
S is the volume of saline to be added.
Once we get the hct where we want, we’ll measure pH and temperature again. At one point we were centrifuging the entire sample to remove the buffy coat layer between the serum and the cells, then mixing it back together but this proved pointless and didn’t really benefit our results or affect our testing any and it was a major pain, so I don’t recommend it.
Once prepared, we can expect the blood to last for five or so hours before it gets questionable. If we’re testing an endovascular device, we’ll pump the blood around a tubing circuit (using a peristaltic pump) and then place the device in the tubing. Preferably the tubing is a similar inner diameter to the artery or vein the device will be used in. You probably want to place the blood reservoir above the test set up and the pump after the test area. Putting the blood reservoir above the test set up ensures a more consistent blood flow. A simple set up is shown below.
We can measure the device performance in blood directly, or we may be interested in something like how much does the device damage the blood, we’ll check the serum and see how red it is in simple terms. If it gets worse over time, then we’re damaging the blood. In this case, for an accurate comparison we need to run a control at the same time. For example, if we have an elaborate pumping system, we’ll run our pump system on one closed circuit and the control (with no device) in another closed circuit and track the hemolysis of both over time.
Aug 29, 2011
Tech Talk – Anatomy Models and Medical Device Testing
One important consideration to take into account while designing and testing medical devices is how they’re going to be used. In the case of vascular devices, they will be used in arteries and veins and I’ll describe some of the test considerations. While performing your verification testing, you want to ensure that you have a reasonable clinical model to perform testing under.
For example you might have a device that is intended to be used in the Right Coronary Artery (RCA) as shown below (image from Wikipedia):
How do you ensure your device works correctly without using it on a person? You find an appropriate model.
Models of various anatomies are available from Elastrat and Shelley Medical Imaging Technologies. ASTM F2394 also contains a schematic of a two dimensional (2D) model recommended by the FDA for some applications. Although it is best to ensure you pick a model based on the vessel characteristics that your device will likely see (such as vessel diameter and tortuosity). Using the ASTM or another off the shelf model can get you into trouble if for example you’re testing a guide catheter and pushing it into the smallest vessels when it is only intended to sit in the aortic arch.
A 2D model is easy to construct and use, but not necessarily the most accurate, it generally consists of just two pieces of machined plastic (bottom with a clear top) with a piece of tubing inserted. The easiest model consists of just a radius as shown below (by the way, I did the diagram myself in MS PowerPoint!).
Ask your clinical source what the tightest radius your device is likely to see in-vivo and simply machine a curve of that diameter. Insert a piece of silicone tubing into that radius to simulate a vessel wall more accurately than hard plastic and you’re ready to get started (you can use pig vessels to line your tortuous pathway, but this is no fun to set up). It is easy to create 2D models of various tortuous pathways. Creganna (Medical Device Technology, May 2006) shows an example tortuous pathway in some of their coronary block model simulated use testing:
You can see the simulated aortic arch as the large looping arch (start from the top most pathway) and then into tighter arteries as the model continues.
A 2D model obviously lacks some realism as it is possible that your device will need to turn multiple planes during use. Although it may be argued that the worst case scenario is actually the 2D model as it stresses only those two dimensions, adding the third dimensions allows additional flexibility in most devices as they are generally concentric. In the case of a non-concentric model, you can perform your testing to the worst case in the 2D model, setting your weakest side up to take the most abuse. The 2D model is also arguably tougher on your device than clinical use as the give is limited to the wall thickness of the tubing you insert.
For 3D models, you’re probably better off buying one from the companies listed above, although they can get expensive, and can be damaged so you need to be careful. An example heart model from Elastrat is shown below.
Once you have your anatomy model, you generally want to further simulate clinical conditions, either by flowing 37C saline through the model, or a mixture of glycerin and water to simulate blood flow.
Now you want to run whatever tests you can using the model. For example, a guide wire or catheter turns to failure test is simple enough, just hold the distal end and turn the proximal end until the device breaks. However, to really simulate use, you should put the device through the model, and while the device is still in the model, hold the distal end then turn the device until it breaks. The addition of the model adds a bit of complexity to the test, but is a better test system.
I have seen the FDA question models, so document your clinical justification why the model you use is appropriate in the protocol. If you’re using a model that you bought, you should ask them to provide you with the justification.
Jul 12, 2011
Tech Talk – Guide Wires
The user requirements of a guide wire are can it quickly get to where it needs to go and then can I reliably deliver devices over it without patient safety issues.
A typical guide wire construction is shown below:
In this diagram, the blue proximal section is the hypotube, similar to a syringe needle (minus the sharp end). The hypotube is generally coated in PTFE- but other coatings are also used. The PTFE coating allows devices to more easily slide over the guide wire and is probably the major component in device “deliverability” on the guide wire side. It is easier to slide a catheter over a guide wire with PTFE than it is to slide a catheter over a guide wire without PTFE. The PTFE coating of the hypotube can be a tricky part of manufacturer and is usually outsourced. Just as your pans vary in quality of the non stick surface, so do guide wires and different variations in processing and materials can affect how well your guide wire delivers devices.
The core wire material is usually nitinol or stainless steel and tapers from the proximal end to the tip. At the tip it is generally flattened. The core wire affects the torquability of the device; you want the tip of your guide wire to turn in a 1:1 ratio with the proximal end. The torquability affects the steerability of the device, can it get to where you want to go. Nitinol core wires are harder to kink, but can lose some torquability. The core wire is attached to the proximal end of the hypotube using solder or adhesive.
In this case an extra wire has been attached to the core wire and the tip to allow better tip shapeability and can give a softer tip. Depending on where you want to go in the vasculature different tip softness is desired, for example in the cerebral vasculature usually the softer the better. Tip softness can be measured as tip load, the amount of force (or load) it requires to bend the tip a certain amount. Obviously you want the tip to flex instead of damage the vessel wall.
More information is available at: PCI equipment: Guidewire selection. And the FDA chimes in with helpful guide wire tips on device safety- always read those IFUs.
Nov 26, 2010
The Attribute Gage R&R
There is not that much information about the attribute gage (gauge) R&R readily available online (that I was able to find), so I thought I'd cover what I've done. First the basics, an attribute is something you can't quantify, generally a visual inspection- is this part "red" or something like that. While it generally seems simple, especially to the technical leads, operators can often get tripped up trying to pass or fail parts based on a description or a couple pictures. In the red example, can operators compare against the Pantone effectively and is this repeatable? You want to have your acceptance criteria and how you are proposing to test it with trained operators set up beforehand.
Your gage R&R should mimic what you do on the line and should be set up that way. If the operator does an inspection on the attribute before passing the part, then a final QC does that same inspection, you have two inspections and this should be part of your testing. You should present this as the entire package when possible. QA types tend to freak out when they hear you would accept a 10% possibility of passing a bad part, when in reality its 0.1%. Having the two (or more) inspections will be really helpful when you get to the acceptance criteria portion.
Test Method: You need to determine the number of operators, parts and trials. Trials is easy, just use three, everyone does, obviously more is better, but three is generally good enough and you don't want to be looking at parts all day. For operators, you need two, but if you have three or more lines or shifts, you can include those easily enough. The number of parts is where it gets tricky and you're going to have to make a judgment call. Generally medical device companies rely on some sort of confidence and reliability based on the severity or RPN of the potential failure, however, in the case of gage R&R everyone seems to follow auto industry guidelines which are usually a smaller quantity, 30 is generally defensible either way.
The Acceptance Criteria: The key one medical device companies are concerned with is the probability of a miss, which is defined as:
- Probability of a Miss = (# times a bad part was passed) / (# of opportunities) [i.e. number of inspections]
- Probability of a Miss (2 inspections) = (# of times a miss by one operator that were missed by another operator) / (# of opportunities)
Other Information: I recommend keeping it simple and only requiring a certain probability of a miss in your protocol (maybe effectiveness as well). You'll want the rest of the information you can collect documented, but that is a business decision, not a safety decision. You can cover:
- Effectiveness - (# of parts correctly identified) / (# of opportunities) [a low effectiveness indicates your process is probably not robust and will give you trouble over time, greater than 70% is generally acceptable]
- Probability of a false alarm = (# times a good part was rejected) / (# of opportunities) [waste]
- Repeatability = (# agreements) / (# parts inspected) [calculate per operator and total, if an operator has low repeatability, less than 80% or so, he or she needs retrained]
- Reproducibility = (# agreements among all operators) / (# parts inspected)
- Bias = (Probability of a false alarm) / (Probability of a miss) [calculate per operator and total]
The Test Setup: For this you'll need two experts in the attribute being inspected, the experts will sort out the good parts from the bad, label them in some fashion, and randomize them. In my experience you should have at least 25% bad parts, even though your process isn't likely to have 25% reject rate (hopefully). It is nice to include some very marginal parts, but those can be hard to find and agree on. Don't have the operators performing the test make the parts if you can avoid it.
The Test: You want to set it up so an operator makes a determination and someone else records it, don't let the operators know the sample being given to them, previous results, or talk amongst themselves. Do the testing on the line and try to keep the production pace. During a gage R&R I find the operators tend to err on the cautious side.
The Results: There you go, you can now say your test method is qualified and have the data to back it up. You can also make operator decisions based on the results, maybe move one around to catch things earlier in the process, which one is the go to person, etc. I find the attribute gage R&R easier to perform than the variable one, yet it is generally more important than a dimensional one from a safety perspective because there are more things I can't measure easily.
Aug 14, 2009
Why do clinical trials in the US?
I'm searching for a reason and here is the one I came up with:
- The world expert resides in the US and you really need her on your team
If you have a standard device, almost anywhere will do. If you have a new device, find a well respected doctor in Europe and see what kind of magic you can work there at half the regulatory burden and less lawsuit liability.
(Photo from Karen Horton)
May 3, 2009
Patching Microsoft Windows on medical devices
Law Firm IT has an interesting post on installing the latest Windows patches on medical devices, including this:
"...because the machines were running an unpatched version of Microsoft's operating system used in embedded devices they were vulnerable.I'm not sure if that is completely correct about a required 90 day notice to the FDA before patching Windows. I'll leave that to a regulatory expert.
Normally, the solution would be simply to install a patch, which Microsoft released in October. But the device manufacturer said rules from the U.S. Food and Drug Administration required that a 90-day notice be given before the machines could be patched."
I do know that any changes to Windows and the medical device software has to be revalidated, at least the potentially affected parts anyway. The revalidation is obviously going to take time and effort and the rewards are often low, especially if you label your device to not be connected to the internet. You want your software guys putting in new features, not screwing around with Windows compatibility for people who are using the device off label. Additionally, the Windows patch needs to be installed, no small feat for a device that is not supposed to be connected to the internet. All of this time adds up and the bottom line is you're never going to be a step ahead of software virii on a medical device, which is why they almost all say do not connect them to the internet.
That being said, Windows with the help of one of several off the shelf software programs, such as Clean Slate, Rollback Rx, and Deep Freeze, can be fairly easily configured so that the chances of a virus are minimized, meaning that you wouldn't have to update Windows with every patch. In fact, this seems like a halfway decent mitigation (along with the aforementioned labeling) to your "device connected to the internet" hazard as part of your risk management.
(Picture by Alexander Fediachov)
Apr 17, 2008
Biocompatibility testing on a formerly radioactive medical device
Patrick in the comments asked a while ago:
We have a radioactive device, and the act of charging with radiation has an impact on the surface chemistry, so we need to conduct biocompat tests. Now, our device has a very rapid decay rate, and with a month or so it is effectively not radioactive any more...but testing house are not willing to test it because of the stigma of having once been radioactive. Any idea how these types of devices should be tested?
I don't understand the device and when exactly it would be used clinically, but I think you'll probably have to find some way to test to ISO 10993. Assuming the device is used clinically after the radiation has dissipated, I'd write up a nice summary report saying as much with relevant radiation references and call every testing lab I could with it. Talk and offer incentives to as many people as possible at each lab until I got a definite no from management, you might be able to get a smaller lab to do it, or even offer to do the extractions yourself or under their supervision at your facility if you have a local lab, non-GLP would be better than nothing. You could try hiring a consultant with good contacts at a testing lab as well, they might be able to persuade the testing lab better than you can. Other than that depending on what device category you fall under maybe you could do the testing yourself, although I admit this is not ideal. I don't have any other ideas beyond those, good luck.
Mar 23, 2008
Moved the company
We moved our happy medical device company down the road and I've been spending my time working on projects related to moving. Thrilling stuff, like changing the address on labeling and other documentation, not to mention tracking down various missing items. The cleanroom was moved and the floor done, electricity hooked up, next the whole cleanroom will be scrubbed, the filters turned on and all the furniture will be cleaned and moved in. Then it will be given a few days before its certified and the bioburden checked.
Once that is done we can get started doing installation qualifications (IQ) on the manufacturing equipment. After that we are planning on redoing any validation that involves a manufacturing procedure, but not revalidating things that involve material properties for example. Since the validations require sterile product made at the new facility, away we go. I've written enough validations the past six months I think I could work them up in my sleep. It actually feels like running in place and I think company morale is low due to lack of cash bonuses.
I have a few other projects I work on when I have time, but it is going to take a while to get back to normal after the move. I can only imagine what a nightmare it would be moving a mid-sized or larger company. Although overall it is a good thing as it means the company is doing well enough to expand and grow.
Labels: labeling, paperwork, product validation
Jan 21, 2008
I am web published!
Medical device blogger addresses technical questions
The picture looked less geeky when I picked it out.
Our packaging design went fairly smoothly and was validated. There are couple cosmetic problems related to workmanship that need ironed out, and a few tweaks to the packaging could help that, but nothing serious. Of course, I blamed all of the problems on manufacturing not following the drawing precisely and QA for not inspecting correctly. Just kidding, I'll make the changes when I have time.
Anyway, I spent about 6 hours editing a software requirements document today and will spend about 6 more hours tomorrow because we thought it would be a good idea to let the software guy rewrite and make pretty the requirements before we submitted to the FDA. Instead of taking the already released document and editing it, he made a brand new one, and I get to reconcile the old with the new, both of which are missing major components. The joys.
Labels: packaging, product validation
Jan 20, 2008
End of the 510(k) line
I know how Creo Quality (sorry to post off of Creo so much, but I can't find many other device blogs) feels, the closer you get to your deadline, the more you think, we should (or think we should) have discovered these problems months ago. Our device incorporates an air detector to ensure air doesn't enter the patient's bloodstream, it is a fairly common accessory for any device with an extracorporeal circuit. We just discovered recently that the software was set up to check the air detector every half second, at our flow rates more than 50 ml of air could have gone by in half a second. Thats is why we do the validations. It is a relatively simple job to change the scan rate and then retest that it works correctly this time, probably takes about a week overall, but at this stage every delay is painful and they start to add up.
I've been fairly lucky this week though and haven't caught much flak, management has been occupied with self inflicted problems that I'm not part of. The 510(k) checklist hasn't had many tasks crossed out the last few weeks though, most of it is due to contractors or consultants having one problem or other. Unfortunately it ripples back and other stuff has to be put on hold. A big piece is due tomorrow (No MLK day break for us!), the final software V&V, we'll see how it looks. Until we have that we haven't been able to finish a lot of traceability documentation and a bit of testing that was waiting for the final software version to be signed off on.
Dec 27, 2007
Bench Testing
I've spent the last few weeks dealing with bench tests as we wait for other results to come in. Bench testing or performance testing gets stuck in section 18 of the 510(k), since it is in the back means they don't read it right? You will save yourself a lot of grief if you read the format guidance and make sure your protocols and reports line up nicely with their requirements, although you will need more than what is listed in the guidance, at least a scope.
For a small company there are several challenges to bench testing, all revolving around the number of people that have enough understanding to run the tests. My company has three plus one consultant that can run the majority of the tests, it is preferable to have employees sign off on everything so the consultant is out, and one of the three has the understanding to run the tests, but not the personality type to see it through. That leaves the two engineers, one of which is on vacation this week, so that leaves me for now. MD&DI sums up the who should do the bench testing very well.
The first problem is the protocol, which must be signed off before the test begins, the problem here is that no one besides the engineer authors are likely to really understand what is going on. This means no problems will be caught until the engineer testers try it for real. Sure, we've tested it some previously, but when everything is recorded things change. I wrote a protocol and discovered I couldn't hold a negative pressure I thought I could so had to change it up a bit. This means rewriting and walking around getting signatures to get it approved before I can start again. This is not much of a problem, unless it is after 3:30pm and QA has gone home for the day. Then I'm forced to wait around until they come in at 9 the next day. I have argued that by having my signature on it that the protocol has therefore been predefined and good to go, but I haven't gained much ground with that.
The next problem is that these tests take time, we are shooting for 24 hours of use. I rallied around testing for 26 hours but my boss vetoed that saying 1.5 times is standard, meaning 36 hour tests and every other day I have to come in at an awkward time (do not worry, I am getting my revenge- see below). I am amazed my wife hasn't accused me of cheating on her yet with the late night stops by work. The 1.5 times the maximum limit you're shooting for is a good rule of thumb, and appropriate here, but it doesn't work for everything, like negative pressures.
The last of my whining centers around the sample sizes that will not be high enough to make everyone happy. With limited product and limited resources, running a dozen 36 hour tests could take a month. Unless you are going to manufacture, sterilize, and shipping simulate a batch of samples yourself in the next week, complaining about sample size doesn't accomplish much. Do a reasonable job and if the FDA picks on it the most likely thing that will happen is they'll ask for more testing.
I mentioned in my previous post that the deadline slipped (still not my fault), this has given me time to come up with some additional bench testing to put in motion. I say put in motion because I was so confident I'd meet my part of the original 510k deadline that I planned a two week Hawaii trip starting one day before the deadline. Now all the loose ends will have to be tied up by my boss and the other engineer, I sorta feel guilty now, but about 20 minutes after landing it will be forgotten. I give the extra testing a 40% chance of not being done when I get back. I have to say though that the last year has been a blast and if you're an engineer with a good work ethic that can tolerate the risk of working for a smaller company then go for it.
Dec 18, 2007
Sterlization finally resolved
I posted previously on failing a natural product sterility (NPS) test in a lot release using ethylene oxide (ETO) sterilization, and we finally have everything somewhat resolved, well really a couple weeks ago we did, but I've been too busy to post. An ISO 11135 lot release requires a half cycle, then biological indicator (BI) and NPS testing which must come back sterile, then a full cycle and further BI testing which comes back sterile. Residuals and pyrogen samples are taken from the full cycle. In the previous post I mentioned we failed the half cycle NPS testing.
After that we made our full cycle the half cycle and tested the full cycle samples for NPS and doubled the sterilization time for the rest of the samples (this still confuses some people at the company). Unfortunately those NPS samples failed as well, or more precisely one of forty samples failed on Day 7 of 14. All BI samples were negative. After much debate and me being the ever eager middleman between management and the sterilizer we decided to double our sterilization time once again then proceed. However, we had to start all over with the samples because we had used so many in testing.
While we were building more samples we looked into the NPS testing. The most notable thing was that we discovered we were using twice as many samples as required, 40 instead of 20. 20 samples assumes you use 10 samples with aerobic media and 10 samples with anaerobic media. You can even use 10 samples if you use the membrane method instead of using the two types of media on the device. For us, each sample was tested individually and had their own large media jar. We made a couple minor, mostly for show changes to how the test was conducted.
So we built more samples and re-sterilized at a fairly ridiculous length of time for the half cycle and twice this ridiculous length of time for the full cycle, the length of time is so high it borders on what the FDA considers non standard sterilization. Hopefully we can lower the time some in an actual sterilization validation in the future. Oddly (IMO), the sterilizer doesn't charge by time in the chamber, so sterilizing for 1 hour costs approximately the same as sterilizing for 5 hours.
It took much begging, but we actually managed to get these sterilized half and full cycle in less than three weeks. We passed all the required testing and we are finally on our way. There is another part to our product sterilization that I'll get back to when its done. Talking to the 510(k) consultant afterwards he said sometimes companies don't file with sterility test results. I'm not exactly sure what they're filing with then.
Since then we've been bench testing like crazy and I have been unsuccessfully trying to delegate tests, but it seems like we don't have any people that understand the product well enough to pass most tests on to. We did have a Dec 31 filing deadline, but its been moved. I am happy to say that my disposables part isn't the thing that is holding everything up despite a completely failed sterilization.
Labels: product validation, sterilization
Nov 1, 2007
Handling samples
As part of our company growing up, and failing a sterility test, we've put the quality department in charge of samples instead of manufacturing. In most companies I've seen, quality generally handles various samples and installs things like biological indicators (spore strips) and other special sample conditions. For us, manufacturing was doing it because the QA/QC department was stretched thin, but now somehow we have acquired more quality personnel than manufacturing personnel. A situation which immobilizes manufacturing at least 2 out of every 5 days and has 50% of them threatening to quit regularly.
Anyway, I mentioned spore strips because I wanted to detail the fun of using them. First off, if you can, use the ones in the glassine envelopes. However, for some reason, they make smaller and different shaped spore indicators, but leave them in the large envelope. This means you may have to use a type that does not come in the envelope or open the envelope and insert the sample into your product. If you own a BI company I see a market opportunity, or maybe I just can't find them and everyone else knows about them.
Because nothing is ever easy, we of course have to use the non-enveloped indicators. My first instinct is that you don't want to do this in the clean room, that seems to be just asking for trouble. (ASFAIK not a problem with the enveloped ones) One of the afore mentioned "quality" personnel will drop one on the floor and then step on it and track it around the room and we'll have to spend the next week disinfecting. Instead, manufacturing is going to bag the product and remove it from the clean room and masked and gloved QA will insert the spore indicators under a fume hood with sterile instruments as needed. They will then rebag the product and seal it immediately outside of the clean room. I ruled out using a clean bench instead of a hood because the air blowing out (instead of up) is more likely to disperse bacteria, most likely right on to the person doing the indicator placement, although I'm not sure how much I should be worried about that. I'm not very happy with this procedure and might rework it on a whim as I think more about it. Moving the product out of the clean room for the prep is likely to add some bioburden to it, although it shouldn't be much and surface bioburden is probably easier to kill, its still better not to have it in the first place.
PS I might sometimes sound mean and sarcastic to readers, but those who know me know I'm only sorta joking and don't take myself too seriously!
Oct 14, 2007
Failed a natural product sterility test
One part of the ETO sterilization validation we're performing involves testing the natural product sterility (NPS). This entails sterilizing the medical device and then immersing the device in a media (you can also run the media through the device). Then removing the media and waiting around to see if it grows anything. We had 40 samples tested, which is standard, I'm not sure if they mix the media together or keep it separate while waiting for stuff to grow. By testing the sterility this way you're ensuring that the product is at least as easy to sterilize as known spore strips. Once your sterilization is fully validated you can skip the NPS testing.
Anyway, our natural product sterility sample failed on day 12 of 14, looks like a do-over with some more aggressive sterilization conditions. Even if it is a false positive (day 12!!), it doesn't matter, its a non-sterile load. Resterilization (or actually now just sterilization) is a tricky mix of old full cycles becoming the new partial cycles and the sample requirements cascade from that.
There is a silver lining, we did pass the bacteriostasis/fungistasis test! Actually, while a bit of a downer we had planned for something like this previously and we have to push some things a bit harder- we need to make up some samples, but overall it is not too bad and our timeline is largely intact. My boss has had these things happen with sterilization before so he had us prepared.
Labels: product validation, sterilization
Oct 5, 2007
Medical device questions asked
I thought I'd answer some questions that people have attempted to find answers for on this blog, keep in mind that I'm no expert- and looking at everything from a USA point of view, you should double check everything I say.
If you validate a medical device with one sterilization method (ETO), do you have to revalidate if you change methods (to gamma)?
Yes, you will almost certainly have to revalidate almost everything. I can't think of any concrete exceptions now, if I do think of some or hear of some I'll add them at a later date. I think you may be able to justify your way out of redoing some performance studies depending on what you had to show.
Do you have to sterilize all medical devices?
No, the term "medical device" covers a wide range of products including things like defibrillators, ECG machines, and external pumps. Some classifications don't require sterilization, find your classification on the FDA website starting here and go from there. A good rule of thumb is if you're going to contact something not normally exposed to outside air, then you have to sterilize it.
What do I need to get a medical device ready for clinical trials?
Medical devices manufactured to GMP and a 510(k) if you're going down a 510(k) path. If you're following a PMA path in my experience you'll need to convince the FDA that you have a quality system and manufacture to GMP, biocompatibility validation, safety validation which may include a few animal studies, and a some decent test results- along with whatever else the FDA and the hospital review board say you need.
What is the accepted standard deviation for medical devices?
I have no idea if this is related to lab results or what, but there is no standard accepted standard deviation for medical devices, however, if you go less than two, you should be prepared to justify it. This does not mean that two is required, you just have to justify it, please keep this in mind if you are transferring from the drug industry!
How can you tell if a medical device needs FDA approval?
Check the FDA web site, it really is very informative. Find your device classification on the FDA website starting here and go from there.
What is a typical day for a medical device representative?
I talk with these guys from time to time, it generally involves driving around to doctor's offices and hospitals to try to sell and then complaining that the company doesn't have any good products in the pipeline.
How many people to hire for a medical device start up?
I think I've covered this before, but I'll go for it again, 1 QA, 1 manufacturing, 1 engineer and 1 management is a starting point to build from. You really do want all of your departments covered from the start or you'll have a big mess on your hands when you try to finalize things.
Are medical devices good or bad?
They are good!
How much money to start a medical device company?
If you have a cheap, simple, 510(k) device and are somewhat knowledgeable about the field, I bet you could do it for less than $1 million. If you have some drug coated, ground breaking, research intensive device you'll need many times that.
What is the medical device standard expiration date?
There is no standard expiration date, but unless there is a good reason otherwise, you can label for 6 whole months without validating the packaging and saying you'll "hand carry" everything to the site. Hand carrying involves someone flying on the plane with the boxes. This is useful because you don't want to validate your packaging until you have a final product. On the other side of things, it used to be you tried to label for a 5 year shelf life, but I think lately a lot of products and companies are content with less, 2 or 3 years is fine.
What is a sample IQ/OQ/PQ?
I think I'll do a post on this later, until then search the MDI archives.
Sep 8, 2007
Moving?
The powers that be have recently started discussing moving the company down the street to a slightly bigger, nicer, cheaper place. If this happens I may have to be restrained from stabbing myself in the eye. While there are certainly good reasons to move a medical device company, the people working on validations never ever want to hear the words "We're moving." To them, this translates into "Everything you've done in the last year is now worthless." Softening the blow with a raise at that point is just good form, if anything a company moves because they're doing well, right?
Moving a medical device manufacturing facility basically means a revalidation of almost everything. Every piece of manufacturing equipment must undergo an installation, operation and performance (or process) qualification (IQ/OQ/PQ- terminology can vary from company to company) when you start to use it. These can range from simple to complex, with most taking less than a day to complete and document. Move a piece of equipment across the room and you are supposed to perform at least the IQ again (which can be very simple) - usually we do all just to be on the safe side. It can become tedious depending on how it is set up, a lefty moving a balance to the other side of the table can give you fits. (Although I don't think I've yet met a QA who kept track of exactly where a piece of equipment was placed, they've relied on Manufacturing to tell them of moves- sounds like something that should be added on the IQ form.) A move to a new building invalidates everything and on top of that you have to show that you can produce product of the same or better quality that you previously produced.
My company and I assume most other companies perform and document these qualifications in house. Engineers generally write them, a tech or engineer will perform them, and then QA will sign off on them and keep them in an equipment log. At my company its a relatively informal process and QA runs the entire show without management involvement.
The silver lining to moving the company is that you don't have revalidate sterility, biocompatibility and maybe some performance testing, but everything else I can think of now is fair game. And of course that raise.
PS: Really nice companies have downloadable sample IQ/OQ/PQs available.
Sep 4, 2007
Validating the packaging process
All sorts of design issues can show up unexpectedly in packaging, a small wrinkle in a bag or pouch seal can mean a complete redesign or worse. My boss's blood pressure skyrockets whenever someone mentions the dye penetration test (ASTM F1929) mentioned in the article. We received a bunch of failed test samples back from a testing lab that were solid coated with dye which supposedly showed how we failed the test, except after the test is over dye coats everything so you can't tell. Ever since then he's hated the test with a passion and if you want to avoid a 15 minute rant its better to stay off the subject.
I was part of a PMA that the FDA specifically required the dye penetration test for to move forward, so you'll likely have to get through it (although you can avoid packaging validation in the early stages of a PMA if you hand carry). The dye penetration test basically consists of pouring a dye solution into the packaging and then observing its progression through the package seal in a certain amount of time. The problem is that the test requires judgment calls regarding wrinkles and has a wide time range. A small wrinkle in the seal can allow the dye to penetrate quickly through the seal.
Dye Penetration Test Results
How do you avoid this problem? There are a couple ways to avoid it completely. If you gamma sterilize you don't need Tyvek breathers and the bags won't leak if you seal them properly in the first place. It is slightly harder to seal completely plastic bags, you can end up with ridiculously large amounts of air in the bag if you're not careful. There is also a company that makes Tyvek windows and you seal the plastic bag, since the Tyvek is not on the edge, you're much less likely to have a problem (I forget the name of the company offhand, but I'll update later). The sterilization degassing time may be longer however. The other packaging option is to use a more expensive tray type instead of a bag.
How do you minimize the problem? Don't handle the packaging bags. It makes me cringe when manufacturing or QA handles the bags roughly. The other thing you can do is properly design the box to limit the movement inside- this is difficult with the larger your product is however.
Aug 28, 2007
Biocompatibility testing of medical devices
I spent today writing the biocompatibility testing protocol which is required for product validation, engineers writing testing protocols is what happens when you don't have dedicated regulatory or QA. I'm sure that is making some of you cringe- but I promise I write a better procedure than blog post.
For the testing, which should be done on the final device after final processing, you basically just round up ISO 10993 and look at the biocompatibility testing chart and then call around for some GLP quotes. The big three places to get biocompatibility testing done seem to be NAMSA, Apptec, and Nelson Labs. There are also regional players, in our neck of the woods its Pacific Biolabs (who have decent information on their webpage here). Most people who have been around for a while have their more liked lab that they use most often. I wouldn't call it a favored lab because the relationship is hot and cold depending on how long the lab is taking- a bit silly since most testing requires standard amounts of extraction, incubation or other times. I annoy them all depending on the whims of my betters, I probably request five quotes for every one test we actually perform, sorry if you're the one doing the quoting, but thats how it works here.
For the most part, our required tests are fairly straight forward, we do have an "Additional test which the FDA may consider applicable" note under one test. In cases like that, you can generally do the test, or spend just as much time and money justifying why you don't have to and then have the FDA tell you they'd really like to see the test done and why don't you just go ahead and do it. We're just going to go ahead and get the additional test done.
I spent a bit of time trying to convince people that Murine Local Lymph Node Assay (LLNA) was the way to go for sensitization testing. Its a newer test, that is quicker and supposedly more accurate, it costs the same, and is also supposedly better from an testing animal welfare standpoint. I made some progress, but we could revert back to the Guinea Pig Maximization test at any time someone gets antsy. We just last year switched to LAL tests from rabbit pyrogen tests.
There really isn't any trouble anticipated, we're using standard tubing, components and assembly methods for the disposable component. These materials have probably been tested thousands of times previously, which is a little depressing, but its less than $20K for the testing for our device which takes about two months, so its not a huge burden for a very nice little piece of mind. The rationale behind the repeated testing is no two processes are exactly the same so better to be safe than sorry. Once the samples get back from sterilization we're good to go and hopefully the remainder of the work required for biocompatibility validation is just sorting out test results into binders.
UPDATE: I just wanted to make clear that not all devices can get away with a relatively easy biocompatibility testing as we can. The longer the device is in contact with the patient, the more rigorous the testing and some of the tests (genotoxocity or implantation for example) can be $20K+ each and some can take 4+ months. I also clarified that the testing should be done on the final device after final processing as per the comments.