Note: This is the final part of the medical devices particle testing tech talk, see part 1 and part 2.
Once the particulate test method has been validated, it is
appropriate to start product testing.
The FDA guidance documents suggest testing finished devices subjected to
sterilization, performing testing on the extremes and an appropriate
intermediate size for the product matrix, and assessing both inter- and
intra-lot variability. A common way to
meet these requirements is to perform testing on samples from design
verification, aging, and three lots of process qualification.
The best practice would be to also test lots produced under
worst case coating process conditions, which is the thickest allowable coating
applied using the minimum cure time, although the FDA did not mention
this. If you do particulate testing as
part of lot release testing, it is in your best interest to test the worst case
coating process conditions.
It is desirable to finish as much testing as much as
possible in one day; this makes the results more consistent and minimizes the
amount of time spent cleaning.
A typical test format is:
- Perform test on water with glassware
- Perform test on water through the model without test device
- Perform test on water through the model after test device is cycled
To test, first ensure the water and glassware to be used are
acceptably clean. For these examples it
is assumed the validated particle method used 50 ml of water. For example, if the test includes using a
syringe to inject water into your model then collecting the effluent in a beaker,
use the syringe to inject 50 ml of water into the sample collection beaker and
test it. The result should show a small
number of particles in the 10+ um bin and very few (i.e. 0, 1 or 2 per ml) in
the larger bins. If necessary, clean
your test glassware some more and then retest.
Next, get baseline results.
This can be done by injecting 50 ml of water through the model, collecting
it in the test collection container, and then performing the particulate
test. This is the baseline and should be
subtracted from your test device results.
Typically, the baseline has more particulates than the glassware test,
but it should still not be that many. If
you see more than 5 large particles (i.e. 50+ um), I would rinse the model with
water and perform the test again. The
baseline test may be performed before every sample test, per sample group, or
per day. Any of these methods is
defensible. You should also re-determine
the baseline if a test condition changes, such as a new bottle of water is
used.
Then you’ll perform your test to typical use
conditions.
As before, a typical test
might be:
a. Fill
model with 10 ml water, collect any effluent in sample container
This step
ensures the model is hydrated prior to use, very few endovascular procedures are
performed with a system that is not hydrated.
If the system is not hydrated the devices will likely generate extra
particulates.
b. Fill
guide catheter with 1 ml water, collect any effluent in sample container
This step
ensures the guide catheter interior is hydrated prior to use, for the same
reasons as listed above.
c. Perform
simulated use with your device which takes 4 ml of water (obviously varies by
device volume), leave device in model, collect any effluent in sample container
This step is
the meat of the test. Simulated use
should match the IFU and typical use.
For example, if you have a guide wire and the IFU states to hydrate it
for 30 seconds, you should hydrate it for 30 seconds prior to insertion into
the RHV, through the catheter and into the model (the water used to hydrate is
not used in the test). Continuing the
guide wire example, the guide wire should be advanced to a clinically relevant
position in the model, and then retracted, the advance and retractions should
be performed a clinically significant number of times. For a PTCA catheter, the FDA guidance
suggests inflating to the maximum labeled diameter.
d. Flush
guide catheter with 10 ml of water, remove device from model, collect any
effluent in sample container
This is a
typical example; the guide catheter is often flushed during endovascular
procedures. Using the guide wire
example, you would flush through the guide catheter because it is standard
practice and you will capture any particles removed from the outside of the guide
wire. Flushing through the guide
catheter ensures you collect the most particles. Alternatively you can perform a flush through
the model with the guide wire in place, but the particles generated by the
guide wire in the guide catheter will not be captured. One could also perform both flushes to be
conservative.
e. Flush
model with 25 ml of water, entirely empty model into sample collection
container
Flushing
after the device is removed from the model ensures that any particles generated
during device removal are captured.
f.
Perform particulate count matching the
validation conditions
Perform the test using the method previously
validated.
g. Flush
the model with water
To ensure the model is clean for the next test, flush
with water. You can determine how much
water is required by testing the effluent after a flushing, or you can perform
a baseline test prior to every test as mentioned above.
h. Identify
particulate (as necessary)
Identifying the type of particulate can be done to
determine the source of the particulates.
It is generally only attempted when an unexpected number of large
particles are detected. TIR42 lists
typical methods for particulate matter determination. To identify the particulate you have to
retain the remainder of the sample, or collect it from the particle counter
effluent. Collecting the sample from the
particle counter effluent can be challenging due to the particle counter
volume.
To analyze your results,
subtract the baseline the sample test results.
If the baseline had a higher result than the test (resulting it a
negative number) it is generally acceptable to change that bin to zero, how to
deal with this situation should be discussed in the protocol. Finally it is generally desirable to convert the
results to a per device basis and determine if the results met the
specification.
This blog is very informative about latest trends about medical devices and news. I thought you can get even interesting news here.
ReplyDeleteOur team is passionate about offering the best solutions in medical field which can be seen through our customer services. For products and services visit:- More Information
ReplyDeleteAny suggestions on how best to prevent and/or clean components toward meeting USP788?
ReplyDeleteWhenever I read Your Post Allways got Something New
ReplyDeleteBaby Care Equipments
Are you a medical equipment dealer? Then you may have probably know the importance of keeping your labs up-to-date with the advanced healthcare products.
ReplyDeleteMedical Equipments | Fetal Monitor | Office Hysteroscopy
This blog is very informative about latest trends about medical devices and news. Now you can shop Online Medical Equipment.
ReplyDeleteI had visited your website which was really good
ReplyDeleteGlassware Manufacture
Really enjoy reading your blog and articles. It's incredible that someone so knowledgeable in the field is willing to put in effort and organize articles like this. Thanks very much! Andrew Wu
ReplyDelete