The FDA guidance (PTCAballoons and stents) points out that particulate matter can be generated by the
manufacturing process or from the breakdown of any coating (e.g., hydrophilic
coating) on the device or from the device packaging. If particles are
introduced in the bloodstream during use, they may present an embolic risk to
the patient. Measurement of the total quantity and size of particulates a
device may generate is an indication of embolic risk.
It used to be that
medical devices didn’t have hydrophilic coatings and particulate really wasn’t
an issue. Then hydrophilic coatings came
along and for a while USP 788 specification was adopted for use with medical
devices. If a small volume injection
could have 6,000 or less particles 10 um (micrometers) or larger and 600
particles 25 um or larger, it seemed reasonable that if a medical device
generated less than that it was okay. A
general note, particle counts are binned as 10+ um, 25+ um, etc. The count for the 25+ um bin is included in
the 10+ um bin, so the particle count will always decrease as the bin size
increases, binning other ways may confuse people.
This worked for a while
until people thought about it more and some of the coatings turned out to
generate large numbers of particles. The
USP 788 specification has a significant issue, there is no discussion of upper
limit of particle sizes, i.e. you could have a bunch of particles half an inch in
diameter and still meet the adopted USP 788 specification. The specification may work for injectables
because you will not get half inch particles in a liquid and certainly can’t
inject them anyway, but with a medical device you just might be able to. Since
that question has come up, AAMI TIR42:2010 w as released with a
section on the clinical significance of particulate matter which basically
concludes that particulates less than 100 um are not a major concern. There is less evidence showing any particles
larger than 100 um are safe (although they may be). This just further highlights the inadequacy
of using USP 788 as a particulate specification for medical devices.
So while you may still
use an adoption of USP 788 as your specification, you do so at your own risk
and you’ll probably need some further explanation for the FDA. What specification you do choose is tricky
though, as you probably do not want to set a specification of zero particles
100 um or larger. First, is this
clinically significant? Second, in my
experience every now and then you will get a particle that large, it may not
even be from your device, but it will show up in your environment results. If you have a history of using USP 788 for
other devices on the market, you can probably use that if you’re comfortable
with it, along with a clinical evaluation of your results on the larger
particles. However, the FDA has left the
door open here for you to accept larger numbers of particles than USP 788, which for some coatings may be required, the AAMI TIR42 standard
references plenty of literature saying large numbers of small particles are
unlikely to do harm. Alternatively, you
could compare your results to results from a similar device on the market, but
this method is riskier, expensive, and is going to be less repeatable.
Part 2 of this series discusses the test method and how to validate it. Also see Part 3, Medical Device Particle Testing.
Very important consideration for designing enhanced functionality in a device. Thanks for pointing this out
ReplyDeleteThank you for your cogent analysis on episat software Brad. It is very informative.
ReplyDeleteCheck this out too::::
Android Development
Ipad Applications development
HIPPA Certified Software
medical software Development
EHR integrations
Top rated blog. FDA or ISO review is an important and conceivably challenging some piece of working together for any medical device maker. For startup medical device makers the first review of your procedures could be as frightening as the first occasion when you safeguarded your strategy for success before investors, submitted your initial 510(k) provision, or conveyed your item to market. Possibly all the more in this way, on the grounds that you realize that ending up on the FDA watch rundown could startle away potential clients and speculators, and maybe extremely incapacitate your diligent work and organization objectives.
ReplyDeleteAll the best~ Hannah
Recently, I've heard about this artificial vision device that helps legally blind people "read" on their own. Although this technology was originally created for visual disability, it does work for people who have reading difficulties, dyslexia or low vision.
ReplyDelete